Skip to article frontmatterSkip to article content
Site not loading correctly?

This may be due to an incorrect BASE_URL configuration. See the MyST Documentation for reference.

AFM

The atomic force microscope maps a surface by touch. A tip a few nanometers sharp, mounted on a flexible cantilever, is scanned across the sample while a feedback loop holds the tip-sample interaction constant; the feedback signal traces out topography with sub-angstrom height resolution. No vacuum, no conductivity requirement, no sample preparation: AFM images metals, oxides, polymers, and wet biological samples equally well, which is why it sits in nearly every materials laboratory and cleanroom. For thin films AFM is the standard measurement of surface roughness, and its quantitative use is the subject of this page.

At a glance
Probe in / signal outSharp tip on a cantilever in mechanical interaction with the surface
InformationTopography, roughness statistics, step heights, grain morphology; mechanical properties
Vertical resolutionBetter than 0.1 nm; atomic steps trivially resolved
Lateral resolution1 to 10 nm, set by tip radius
Field of viewUp to about 100 um; scan times of minutes
Sample requirementsAny reasonably smooth solid, in air, liquid, or vacuum; nondestructive

The transducer

The cantilever is a calibrated spring: forces between tip and sample of piconewtons to nanonewtons deflect a cantilever of stiffness kk from about 0.01 N/m (soft contact levers) to about 40 N/m (stiff tapping levers), and the deflection is read out by a laser reflected off the cantilever’s back onto a split photodiode. The optical lever multiplies angular deflection by the beam path length, converting sub-angstrom motions into measurable photocurrent differences: mechanical amplification with no contact and negligible noise, the enabling trick of practical AFM. Oscillating modes add the cantilever’s resonance, f0=(1/2π)k/mefff_0 = (1/2\pi)\sqrt{k/m_{\mathrm{eff}}}, typically tens to hundreds of kHz, whose amplitude, phase, and frequency become measurement channels. Quantitative work calibrates both conversion factors rather than trusting nominal values: the deflection sensitivity from a force curve on a hard surface, and the spring constant most conveniently from the cantilever’s thermal vibration spectrum (the equipartition theorem turns the measured Brownian motion into kk), the standard “thermal tune.”

A force-distance curve, deflection versus tip approach, is the technique’s central calibration and diagnostic tool: approaching from far, the tip feels growing van der Waals attraction until the force gradient exceeds kk and the tip snaps into contact; pressing further climbs the repulsive wall; retracting shows adhesion hysteresis before snap-off, with the pull-off force measuring tip-sample adhesion (interpreted through contact mechanics models when quantitative adhesion or modulus is the goal). Everything about imaging modes follows from where on this curve the instrument operates.

Imaging modes

AFM modes on the force curve

Module 7 in one figure. Every scanning probe mode is a choice of what to hold constant on this curve, or of what second quantity to record alongside the height.

Quantitative roughness metrology

The AFM’s headline number for thin films is roughness, and using it well takes more care than reading the RMS value off the screen. The standard statistics, average roughness RaR_a and RMS roughness RqR_q (the standard deviation of heights), characterize the height distribution only; two surfaces with identical RqR_q can be morphologically different, one gently undulating, one sharply jagged. The power spectral density (PSD) of the height map resolves roughness by lateral wavelength, separating measurement noise, grain-scale texture, and long-range waviness, and is the right basis for comparing films or feeding roughness into optical and XRR models. Any reported roughness is a bandwidth-limited quantity: the scan size sets the longest wavelength included and the pixel spacing the shortest, so RqR_q grows with scan size on most real surfaces, and a roughness value reported without its scan size is incomplete. Comparing an AFM RqR_q against an XRR roughness, which averages a millimeter footprint with its own bandwidth weighting, requires exactly this thinking, and agreement within tens of percent is success.

Step-height measurement is the other precision use: an AFM profile across a masked or etched step measures film thickness absolutely, to sub-nanometer accuracy, with no optical model at all, which is why step-height standards calibrate the instrument’s z axis and why a deliberately patterned step is the fastest arbiter when ellipsometry and XRR disagree.

roughness against measurement bandwidth

One surface measured two ways. The scan size sets the longest wavelength included and the pixel size the shortest, so RqR_q without both numbers is not a measurement.

Artifacts

Tip shape simulation

Every AFM image is the interaction of two shapes: the surface and the tip. The demonstration below scans a realistic tip, an apex sphere blended into a conical shank, across known test structures, drawn at true aspect ratio so the shapes are not exaggerated. The tip presets carry representative radii and sidewall angles: a standard Si probe, a very sharp Si3_3N4_4 probe, and a blunt diamond-coated probe for conductive or wear-heavy work. Sweep the radius on the particles preset and watch every feature broaden as roughly 2rR2\sqrt{rR} while heights stay correct; try the trench, where the sidewall angle, not the radius, decides whether the floor is reachable; and enable the double tip for the characteristic ghost-image doubling artifact. The measured RqR_q is always at or below the true value, because the tip is a low-pass filter.

Other artifacts

Beyond tip shape, the standard artifact checklist: feedback ringing on steep edges (visible as overshoot on one side), thermal drift and piezo creep skewing slow scans, scanner bow adding false long-wavelength curvature, and the line-by-line leveling that every AFM image receives, which is necessary (each scan line has an arbitrary offset) but can erase real long-wavelength structure or create streaks from particles dragged by the tip. The working rules are simple: know the tip condition (image a known sharp standard when in doubt), image the same area twice with the scan rotated, and treat any feature at the resolution limit with suspicion.

Computed force-distance curve showing snap-in on approach, adhesion hysteresis on retract, and the repulsive contact regime

The force-distance curve. A computed cycle (van der Waals attraction plus contact repulsion against a 0.6 N/m lever): snap-in where the force gradient exceeds the spring constant, the repulsive regime used for imaging setpoints, and the pull-off event that measures adhesion.

References and further reading

  1. B. Voigtländer, Atomic Force Microscopy, 2nd ed., Springer (2019). The modern instrument-level treatment.

  2. F. J. Giessibl, Advances in atomic force microscopy, Reviews of Modern Physics 75, 949 (2003). Giessibl (2003)

  3. G. Binnig, C. F. Quate, and Ch. Gerber, Atomic force microscope, Physical Review Letters 56, 930 (1986). Binnig et al. (1986)

References
  1. Giessibl, F. J. (2003). Advances in atomic force microscopy. Reviews of Modern Physics, 75(3), 949–983. 10.1103/revmodphys.75.949
  2. Binnig, G., Quate, C. F., & Gerber, Ch. (1986). Atomic Force Microscope. Physical Review Letters, 56(9), 930–933. 10.1103/physrevlett.56.930